eurqclog.20050829.000000.txt Data Quality Summary for Eureka SEARCH Research Station Summary created: Tue Aug 30 00:31:12 2005 Value Interpretation -1: No assessment preformed. 0: All values out of range. 1: Some values in and out of range. 2: All data present and all values in range. ----------------------------------------------------------- STATUS FLAGS ----------------------------------------------------------- MMCR: Millimeter Cloud Radar Instrument Status: 1 Data Continuity: 2 Data Quality: -1 MWR: Microwave Radiometer Instrument Status: -1 Data Continuity: -1 Data Quality: -1 AHSRL: Arctic High Spectral Resolution Lidar Instrument Status: -1 Data Continuity: 1 Data Quality: -1 DMS: Data Management System Filesystem Size Used Avail Use% Mounted on /dev/sda1 2.2T 140G 2.0T 7% /data /dev/sdd1 230G 93M 218G 1% /media/usbdisk /dev/hda1 4.0G 307M 3.5G 9% / ------------------------------------------------------------ --------------- ERROR SUMMARIES --------------- MMCR: Instrument Status: 1 20052410100Health.txt: temperature (indoor): 17/28 out of range ( 5.00 to 25.00), Mean: 25.53, StdDev: 1.48 20052410100Health.txt: UPSBatteryVoltage: 115/115 out of range ( 13.25 to 14.50), Mean: 12.02, StdDev: 0.02 20052410200Health.txt: temperature (indoor): 29/29 out of range ( 5.00 to 25.00), Mean: 28.95, StdDev: 0.85 20052410200Health.txt: UPSBatteryVoltage: 116/116 out of range ( 13.25 to 14.50), Mean: 12.05, StdDev: 0.02 20052410300Health.txt: temperature (indoor): 29/29 out of range ( 5.00 to 25.00), Mean: 31.38, StdDev: 0.62 20052410300Health.txt: UPSBatteryVoltage: 115/115 out of range ( 13.25 to 14.50), Mean: 12.02, StdDev: 0.01 20052410400Health.txt: temperature (indoor): 29/29 out of range ( 5.00 to 25.00), Mean: 32.61, StdDev: 0.50 20052410400Health.txt: UPSBatteryVoltage: 116/116 out of range ( 13.25 to 14.50), Mean: 12.05, StdDev: 0.02 20052410500Health.txt: temperature (indoor): 29/29 out of range ( 5.00 to 25.00), Mean: 33.46, StdDev: 0.70 20052410500Health.txt: UPSBatteryVoltage: 116/116 out of range ( 13.25 to 14.50), Mean: 12.04, StdDev: 0.02 20052410600Health.txt: temperature (indoor): 29/29 out of range ( 5.00 to 25.00), Mean: 34.13, StdDev: 0.48 20052410600Health.txt: UPSBatteryVoltage: 116/116 out of range ( 13.25 to 14.50), Mean: 12.02, StdDev: 0.01 20052410700Health.txt: temperature (indoor): 29/29 out of range ( 5.00 to 25.00), Mean: 34.53, StdDev: 0.36 20052410700Health.txt: UPSBatteryVoltage: 116/116 out of range ( 13.25 to 14.50), Mean: 12.01, StdDev: 0.01 20052410800Health.txt: temperature (indoor): 29/29 out of range ( 5.00 to 25.00), Mean: 35.04, StdDev: 0.48 20052410800Health.txt: UPSBatteryVoltage: 115/115 out of range ( 13.25 to 14.50), Mean: 12.00, StdDev: 0.01 20052410900Health.txt: temperature (indoor): 29/29 out of range ( 5.00 to 25.00), Mean: 35.54, StdDev: 0.41 20052410900Health.txt: UPSBatteryVoltage: 116/116 out of range ( 13.25 to 14.50), Mean: 12.02, StdDev: 0.02 20052411000Health.txt: temperature (indoor): 29/29 out of range ( 5.00 to 25.00), Mean: 35.78, StdDev: 0.42 20052411000Health.txt: UPSBatteryVoltage: 115/115 out of range ( 13.25 to 14.50), Mean: 12.01, StdDev: 0.01 20052411100Health.txt: temperature (indoor): 29/29 out of range ( 5.00 to 25.00), Mean: 35.74, StdDev: 0.53 20052411100Health.txt: UPSBatteryVoltage: 116/116 out of range ( 13.25 to 14.50), Mean: 12.03, StdDev: 0.02 20052411200Health.txt: temperature (indoor): 29/29 out of range ( 5.00 to 25.00), Mean: 35.68, StdDev: 0.52 20052411200Health.txt: UPSBatteryVoltage: 116/116 out of range ( 13.25 to 14.50), Mean: 12.02, StdDev: 0.01 20052411300Health.txt: temperature (indoor): 29/29 out of range ( 5.00 to 25.00), Mean: 35.58, StdDev: 0.35 20052411300Health.txt: UPSBatteryVoltage: 115/115 out of range ( 13.25 to 14.50), Mean: 12.02, StdDev: 0.02 20052411400Health.txt: temperature (indoor): 25/28 out of range ( 5.00 to 25.00), Mean: 31.74, StdDev: 4.16 20052411400Health.txt: humidity (indoor): 10/28 out of range ( 10.00 to 75.00), Mean: 11.37, StdDev: 4.48 20052411400Health.txt: UPSBatteryVoltage: 115/115 out of range ( 13.25 to 14.50), Mean: 12.04, StdDev: 0.02 20052411500Health.txt: UPSBatteryVoltage: 116/116 out of range ( 13.25 to 14.50), Mean: 12.00, StdDev: 0.02 20052411600Health.txt: UPSBatteryVoltage: 116/116 out of range ( 13.25 to 14.50), Mean: 12.02, StdDev: 0.02 20052411700Health.txt: UPSBatteryVoltage: 115/115 out of range ( 13.25 to 14.50), Mean: 11.99, StdDev: 0.03 20052411800Health.txt: UPSBatteryVoltage: 116/116 out of range ( 13.25 to 14.50), Mean: 11.96, StdDev: 0.03 20052411900Health.txt: UPSBatteryVoltage: 116/116 out of range ( 13.25 to 14.50), Mean: 11.92, StdDev: 0.01 20052412100Health.txt: UPSBatteryVoltage: 115/115 out of range ( 13.25 to 14.50), Mean: 11.93, StdDev: 0.03 20052412200Health.txt: UPSBatteryVoltage: 116/116 out of range ( 13.25 to 14.50), Mean: 11.99, StdDev: 0.06 20052412300Health.txt: UPSBatteryVoltage: 116/116 out of range ( 13.25 to 14.50), Mean: 12.02, StdDev: 0.04 AHSRL: Data Continuity: 1 Data recovery: 93.7% DMS: RAID Status from dmesg 3w-9xxx: scsi0: Found a 3ware 9000 Storage Controller at 0xfb200000, IRQ: 201. 3w-9xxx: scsi0: Firmware FE9X 2.04.00.005, BIOS BE9X 2.03.01.047, Ports: 12. 3w-9xxx: scsi0: ERROR: (0x03:0x0104): SGL entry has illegal length:address=0x3678C000, length=0xFF, cmd=X. ---------------------------------------------------------