09-sep-11 21:19:54 SERIOUS INSTRUMENT PROBLEM !!! Reference beam OK, Sensing beam low on lamp 3 Tape Advance mechanism problem --- 09-sep-11 21:19:55 Measurements ended --- . Number of lamps (L) = 7 . Filter running time = 59.33 hours . Total airflow this filter = 17.54 cubic meters . Mean BC concentration of all lamps = 32 ng/vm3 . Report for lamp 1 (370 nm): . Ending voltages: sen(beam) = .132 sen(zero) = 0.021 . ref(beam) = 1.512 ref(zero) = 0.020 . Optical attenuation of filter deposit = 42 . Total aerosol black carbon on filter = 534 ng Mean aerosol black carbon concentration = 30 ng/vm3 --- The reference beam showed lamp intensity fluctuations of 79 ppm. --- Report for lamp 2 (470 nm): . Ending voltages: sen(beam) = .695 sen(zero) = 0.021 . ref(beam) = 1.677 ref(zero) = 0.020 . Optical attenuation of filter deposit = 33 . Total aerosol black carbon on filter = 537 ng Mean aerosol black carbon concentration = 30 ng/vm3 --- The reference beam showed lamp intensity fluctuations of 25 ppm. --- Report for lamp 3 (520 nm): . Ending voltages: sen(beam) = .121 sen(zero) = 0.021 . ref(beam) = 2.108 ref(zero) = 0.020 . Optical attenuation of filter deposit = 30 . Total aerosol black carbon on filter = 535 ng Mean aerosol black carbon concentration = 30 ng/vm3 --- The reference beam showed lamp intensity fluctuations of 22 ppm. --- Report for lamp 4 (590 nm): . Ending voltages: sen(beam) = .176 sen(zero) = 0.021 . ref(beam) = 3.226 ref(zero) = 0.020 . Optical attenuation of filter deposit = 27 . Total aerosol black carbon on filter = 560 ng Mean aerosol black carbon concentration = 32 ng/vm3 --- The reference beam showed lamp intensity fluctuations of 76 ppm. --- Report for lamp 5 (660 nm): . Ending voltages: sen(beam) = 1.113 sen(zero) = 0.021 . ref(beam) = 2.403 ref(zero) = 0.020 . Optical attenuation of filter deposit = 25 . Total aerosol black carbon on filter = 583 ng Mean aerosol black carbon concentration = 33 ng/vm3 --- The reference beam showed lamp intensity fluctuations of 37 ppm. --- Report for lamp 6 (880 nm): . Ending voltages: sen(beam) = .346 sen(zero) = 0.021 . ref(beam) = 2.579 ref(zero) = 0.020 . Optical attenuation of filter deposit = 19 . Total aerosol black carbon on filter = 594 ng Mean aerosol black carbon concentration = 33 ng/vm3 --- The reference beam showed lamp intensity fluctuations of 54 ppm. --- Report for lamp 7 (950 nm): . Ending voltages: sen(beam) = .635 sen(zero) = 0.021 . ref(beam) = 2.517 ref(zero) = 0.020 . Optical attenuation of filter deposit = 17 . Total aerosol black carbon on filter = 583 ng Mean aerosol black carbon concentration = 33 ng/vm3 --- The reference beam showed lamp intensity fluctuations of 58 ppm. --- Disk space = 2.514534E+08 bytes free: One entry at every 5 minutes. There is space for 53455 hours ( 2227 days) more data --- Disk space = 2.511585E+08 bytes free: One entry at every 5 minutes. There is space for 53393 hours ( 2225 days) more data --- 09-sep-11 21:36:59 Tape feeder mechanism advancing for 1 spot(s). --- 09-sep-11 21:40:37 ============================================================= Automatic re-start ============================================================= --- MAGEE SCIEN. AETHALOMETER --- Aethalometer No. 774 --- 09-sep-11 21:44:56 Measurements started Starting voltages: . sen(zero) = 0.021 ref(zero) = 0.020 . Lamp 1 (370 nm): sen(beam) = .204 ref(beam) = 1.206 . Lamp 2 (470 nm): sen(beam) = .695 ref(beam) = 1.554 . Lamp 3 (520 nm): sen(beam) = .363 ref(beam) = 1.880 . Lamp 4 (590 nm): sen(beam) = .291 ref(beam) = 3.063 . Lamp 5 (660 nm): sen(beam) = .954 ref(beam) = 2.435 . Lamp 6 (880 nm): sen(beam) = .168 ref(beam) = 2.443 . Lamp 7 (950 nm): sen(beam) = .278 ref(beam) = 2.446 ---