Nephelometer Configuration Data Analog Output Range: Zero V Full Scale Relative Humidity -> 0 100 Barometric Pressure -> 0 1200 Sample Temperature -> 0 400 Scattering -> 7 2 (Offset=1.00000E-7) Channel Type Channel Type 0 9 (Relative Humidity) 5 5 (Red õsp) 1 1 (Barometric Pressure) 6 6 (Blue õbsp) 2 2 (Sample Temperature) 7 7 (Green õbsp) 3 3 (Blue õsp) 8 8 (Red õbsp) 4 4 (Green õsp) Calibration Points: Low Bits Low Value High Bits High Value Relative Humidity -> 250 120 645 968 Barometric Pressure-> 15359 1266 55291 9944 Sample Temperature -> 0 0 39175 2979 Inlet Temperature -> 0 0 38991 2979 Calibration Constants: K1 K2 K3 K4 Blue -> 20000 2.648E-3 2.789E-5 5.180E-1 Green-> 20000 2.767E-3 1.226E-5 5.180E-1 Red -> 20000 2.005E-3 4.605E-6 5.090E-1 Data Delimiter: Comma Calibration Label: TiksiCal_JB_1_5.6.2018_after_service Backscatter Shutter Mode: Enabled Autozero Baseline Measurement: Normal Mode Lamp Power (Watts) : 75 Auxiliary BNC Output (millivolts): 0 Time for Averaging (sec.) : 300 Time to Blank Valve (sec.) : 150 Time between Autozeros (sec.) : 3300 Time for AutoZero Measurement (sec.): 300 PMT Voltage: Blue = 950 Green = 1000 Red = 1000 Zero Baseline Values: Total Scatter BackScatter Rayleigh Scatter Blue -> 2.847E-5 1.361E-5 1.208E-3 Green -> 1.387E-5 6.119E-6 5.312E-4 Red -> 9.942E-6 5.532E-6 1.995E-4