PSA file: C:\DATA\CTDData\2019\2019_config\SE-19-04.psa Date: 06/23/2019 Instrument configuration file: C:\DATA\CTDData\2019\SE-19-04\SE-19-04_testStation.xmlcon Configuration report for SBE 911plus/917plus CTD ------------------------------------------------ Frequency channels suppressed : 0 Voltage words suppressed : 0 Computer interface : RS-232C Deck unit : SBE11plus Firmware Version >= 5.0 Scans to average : 1 NMEA position data added : Yes NMEA depth data added : No NMEA time added : Yes NMEA device connected to : PC Surface PAR voltage added : No Scan time added : No 1) Frequency 0, Temperature Serial number : 2779 Calibrated on : 12-Dec-18 G : 4.37483153e-003 H : 6.48519335e-004 I : 2.35064839e-005 J : 2.17701366e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 2) Frequency 1, Conductivity Serial number : 2326 Calibrated on : 12-Dec-18 G : -1.00830993e+001 H : 1.47170559e+000 I : -1.10364747e-004 J : 9.21219221e-005 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 3) Frequency 2, Pressure, Digiquartz with TC Serial number : 91660 Calibrated on : 13-Feb-19 C1 : -4.100246e+004 C2 : 3.759536e-001 C3 : 1.282219e-002 D1 : 3.715184e-002 D2 : 0.000000e+000 T1 : 2.926839e+001 T2 : -2.159682e-004 T3 : 4.670178e-006 T4 : 0.000000e+000 T5 : 0.000000e+000 Slope : 0.99955355 Offset : -5.29628 AD590M : 1.128000e-002 AD590B : -8.589280e+000 4) Frequency 3, Temperature, 2 Serial number : 1556 Calibrated on : 12-Dec-18 G : 4.84648358e-003 H : 6.80870698e-004 I : 2.66130579e-005 J : 2.07001603e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 5) Frequency 4, Conductivity, 2 Serial number : 3461 Calibrated on : 12-Dec-18 G : -1.00417448e+001 H : 1.54557506e+000 I : -2.45134692e-003 J : 2.76679565e-004 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 6) A/D voltage 0, Oxygen, SBE 43 Serial number : 0715 Calibrated on : 03-Jan-19 Equation : Sea-Bird Soc : 5.72000e-001 Offset : -4.85800e-001 A : -4.12390e-003 B : 1.85860e-004 C : -3.04600e-006 E : 3.60000e-002 Tau20 : 1.18000e+000 D1 : 1.92634e-004 D2 : -4.64803e-002 H1 : -3.30000e-002 H2 : 5.00000e+003 H3 : 1.45000e+003 7) A/D voltage 1, Free 8) A/D voltage 2, Oxygen, SBE 43, 2 Serial number : 0593 Calibrated on : 03-Jan-19 Equation : Sea-Bird Soc : 4.67810e-001 Offset : -4.98100e-001 A : -4.07360e-003 B : 2.19980e-004 C : -3.35190e-006 E : 3.60000e-002 Tau20 : 9.90000e-001 D1 : 1.92634e-004 D2 : -4.64803e-002 H1 : -3.30000e-002 H2 : 5.00000e+003 H3 : 1.45000e+003 9) A/D voltage 3, Free 10) A/D voltage 4, Fluorometer, WET Labs ECO-AFL/FL Serial number : 4108 Calibrated on : 15 Jan 2019 Dark output : 0.0500 Scale factor : 1.00000000e+001 11) A/D voltage 5, Free 12) A/D voltage 6, Altimeter Serial number : 968.8019 Calibrated on : 01/01/2000 Scale factor : 15.000 Offset : -0.500 13) A/D voltage 7, Free Scan length : 41 --------------------------------------------- Pump Control This setting is only applicable to a custom build of the SBE 9plus. Enable pump on / pump off commands: NO --------------------------------------------- Data Acquisition: Archive data: YES Delay archiving: NO Data archive: C:\DATA\CTDData\2019\SE-19-04\deck tests\2740_port.hex Timeout (seconds) at startup: 20 Timeout (seconds) between scans: 20 --------------------------------------------- Instrument port configuration: Port = COM1 Baud rate = 19200 Parity = N Data bits = 8 Stop bits = 1 --------------------------------------------- Water Sampler Data: Water Sampler Type: None --------------------------------------------- Header information: Header Choice = Prompt for Header Information prompt 0 = Cruise prompt 1 = Project Title prompt 2 = Chief Scientist prompt 3 = Survey Tech prompt 4 = Date prompt 5 = Station prompt 6 = Depth --------------------------------------------- TCP/IP - port numbers: Data acquisition: Data port: 49163 Status port: 49165 Command port: 49164 Remote bottle firing: Command port: 49167 Status port: 49168 Remote data publishing: Converted data port: 49161 Raw data port: 49160 --------------------------------------------- Miscellaneous data for calculations Depth, Average Sound Velocity, and TEOS-10 Latitude when NMEA is not available: 19.000 Longitude when NMEA is not available: 0.000 Average Sound Velocity Minimum pressure [db]: 20.000 Minimum salinity [psu]: 20.000 Pressure window size [db]: 20.000 Time window size [s]: 60.000 Descent and Acceleration Window size [s]: 2.000 Plume Anomaly Theta-B: 0.000 Salinity-B 0.000 Theta-Z / Salinity-Z 0.000 Reference pressure [db] 0.000 Oxygen Window size [s]: 2.000 Apply hysteresis correction: 1 Apply Tau correction: 1 Potential Temperature Anomaly A0: 0.000 A1: 0.000 A1 Multiplier: Salinity --------------------------------------------- Serial Data Output: Output data to serial port: YES Seconds between updates: 0.000 Port = COM7 Baud rate = 9600 Parity = N Data bits = 8 Stop bits = 1 Variables: Digits Variable Name [units] ------ --------------------- 4 Descent Rate [m/s] 4 Pressure, Digiquartz [db] 4 Temperature [ITS-90, deg C] 4 Salinity, Practical [PSU] 7 Conductivity [S/m] 4 Fluorescence, WET Labs ECO-AFL/FL [mg/m^3] --------------------------------------------- Mark Variables: Variables: Digits Variable Name [units] ------ --------------------- 0 Scan Count 4 Pressure, Digiquartz [db] 4 Depth [salt water, m] 5 Temperature [ITS-90, deg C] 7 Conductivity [S/m] 5 Salinity, Practical [PSU] 0 Bottles Fired 0 Bottle Position in Carousel --------------------------------------------- Shared File Output: Output data to shared file: NO --------------------------------------------- TCP/IP Output: Raw data: Output raw data to socket: NO XML wrapper and settings: NO Seconds between raw data updates: 0.000 Converted data: Output converted data to socket: NO XML format: NO --------------------------------------------- SBE 11plus Deck Unit Alarms Enable minimum pressure alarm: NO Enable maximum pressure alarm: NO Enable altimeter alarm: NO --------------------------------------------- SBE 14 Remote Display Enable SBE 14 Remote Display: NO --------------------------------------------- PC Alarms Enable minimum pressure alarm: NO Enable maximum pressure alarm: NO Enable altimeter alarm: NO Enable bottom contact alarm: NO Alarm uses PC sound card. --------------------------------------------- Options: Prompt to save program setup changes: YES Automatically save program setup changes on exit: NO Confirm instrument configuration change: YES Confirm display setup changes: YES Confirm output file overwrite: YES Check scan length: YES Compare serial numbers: YES Maximized plot may cover Seasave: NO